TopicId216
Errors related to the PoData device are viewed in the form of text that usually begins with:
PoData Error n.
followed by the error n° and possibly a description and other additional information.
For example, if the PoData device is disconnected from the PC, the text displayed would be as follows:
PoData Error n.110 - PoData USB device not found
An error table is provided below, with a description of the error and possible solutions for the same next to it.
Error n° |
Description |
Solution |
5 |
Unable to execute xxx command, where xxx is the command code. This error means that the PS2 board is unable to execute a command due to problems on the board itself or the conditions of use. For example, the Clear Offset command, with a weight acting on a cell, could give rise to this error. |
Contact the Technical support. |
104 |
USB reading time out - the maximum time for receiving a reply from the USB port has expired. |
See the following chapter: PoData interfacing errors |
105 |
Error in reading the firmware version |
See the following chapter: PoData interfacing errors |
110 |
The PoData device is disconnected from the PC or the PC is unable to communicate with the PoData. |
Connect the PoData to the PC, if the error persists contact the Technical support. |
209 |
IO error on the PoData device |
See the following chapter: PoData interfacing errors |
PoData interfacing errors
Some errors cause interfacing problems between the GPS software and the PoData device.
Communication between the GPS software and the PoData device takes place via various elements:
In general, each element involved must be good quality and must be installed correctly.
When a USB HUB is used, it must have an electric power feed.
If interface errors occur, the functionality of the elements involved must be checked. This check must only be done by expert, competent personnel. For further information contact the Technical support.